DocumentCode :
1211897
Title :
A low-power low-noise CMOS amplifier for neural recording applications
Author :
Harrison, Reid R. ; Charles, Cameron
Author_Institution :
Dept. of Electr. & Comput. Eng., Univ. of Utah, Salt Lake City, UT, USA
Volume :
38
Issue :
6
fYear :
2003
fDate :
6/1/2003 12:00:00 AM
Firstpage :
958
Lastpage :
965
Abstract :
There is a need among scientists and clinicians for low-noise low-power biosignal amplifiers capable of amplifying signals in the millihertz-to-kilohertz range while rejecting large dc offsets generated at the electrode-tissue interface. The advent of fully implantable multielectrode arrays has created the need for fully integrated micropower amplifiers. We designed and tested a novel bioamplifier that uses a MOS-bipolar pseudoresistor element to amplify low-frequency signals down to the millihertz range while rejecting large dc offsets. We derive the theoretical noise-power tradeoff limit - the noise efficiency factor - for this amplifier and demonstrate that our VLSI implementation approaches this limit by selectively operating MOS transistors in either weak or strong inversion. The resulting amplifier, built in a standard 1.5-μm CMOS process, passes signals from 0.025Hz to 7.2 kHz with an input-referred noise of 2.2 μVrms and a power dissipation of 80 μW while consuming 0.16 mm2 of chip area. Our design technique was also used to develop an electroencephalogram amplifier having a bandwidth of 30 Hz and a power dissipation of 0.9 μW while maintaining a similar noise-power tradeoff.
Keywords :
CMOS analogue integrated circuits; biomedical electronics; electroencephalography; integrated circuit design; integrated circuit noise; low-power electronics; neurophysiology; operational amplifiers; preamplifiers; prosthetics; 0.025 Hz to 7.2 kHz; 0.9 muW; 1.5 micron; 30 Hz; 80 muW; CMOS process; MOS transistor selective operation; MOS-bipolar pseudoresistor element; VLSI implementation; bioamplifier; dc offset rejection; electrode-tissue interface; electroencephalogram amplifier; fully implantable multielectrode arrays; fully integrated micropower amplifiers; input-referred noise; low-frequency signal amplification; low-noise low-power biosignal amplifiers; low-power low-noise CMOS amplifier; millihertz-to-kilohertz range; neural recording applications; noise efficiency factor; noise-power tradeoff limit; power dissipation; strong inversion; subthreshold circuit design; weak inversion; DC generators; Low-frequency noise; Low-noise amplifiers; MOSFETs; Power amplifiers; Power dissipation; Signal design; Signal generators; Testing; Very large scale integration;
fLanguage :
English
Journal_Title :
Solid-State Circuits, IEEE Journal of
Publisher :
ieee
ISSN :
0018-9200
Type :
jour
DOI :
10.1109/JSSC.2003.811979
Filename :
1201998
Link To Document :
بازگشت