DocumentCode :
1212161
Title :
Development of a cesium atomic fountain frequency standard
Author :
Kwon, Taeg Yong ; Lee, Ho Seong ; Yang, Sung Hoon ; Park, Sang Eon
Author_Institution :
Time & Frequency Lab., Korea Res. Inst. of Stand. & Sci., Daejeon, South Korea
Volume :
52
Issue :
2
fYear :
2003
fDate :
4/1/2003 12:00:00 AM
Firstpage :
263
Lastpage :
266
Abstract :
We report on the development of a cesium atomic fountain frequency standard at the Korea Research Institute of Standards and Science (KRISS). Cesium atoms were launched upwards by using an optical moving molasses method after they were cooled in a magnetooptical trap. The temperature of launched atoms was about 2.5 μK. From our cesium atomic fountain, we were able to observe Ramsey fringes of 1 Hz linewidth on the 9.2 GHz clock transition of cesium atoms.
Keywords :
caesium; fluorescence; frequency standards; laser cooling; magneto-optical effects; 1 Hz; 2.5 muK; 9.2 GHz; Cs; Ramsey fringe linewidth; atomic clock transitions; cesium atom launching; cesium atomic fountain frequency standards; fluorescence; laser cooled atoms; laser cooling; launched atom temperature; magnetooptical traps; optical moving molasses methods; Atom lasers; Atom optics; Atomic beams; Charge carrier processes; Coils; Fluorescence; Frequency; Laboratories; Magnetic shielding; Standards development;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2003.811680
Filename :
1202025
Link To Document :
بازگشت