Title :
Attenuation measurement system in the frequency range of 10 to 100 MHz
Author :
Widarta, Anton ; Tomoteru, Kawakami
Author_Institution :
Nat. Metrol. Inst. of Japan, Tsukuba, Japan
fDate :
4/1/2003 12:00:00 AM
Abstract :
An attenuation measurement system, which employs a traceable inductive voltage divider (IVD) at 1 kHz as a reference standard and dual-channel intermediate frequency (IF) substitution method, is developed as an attenuation standard in the frequency range of 10 to 100 MHz. High isolation between the channels is achieved by introducing an assembly of light emitting diode, optical fiber, and photodetector into the system. Sources of uncertainties of the system, such as leakage, are discussed and the expanded uncertainties are expected to be 0.001 dB for an attenuation range up to 20 dB and 0.04 dB for 80 dB.
Keywords :
attenuation measurement; measurement standards; measurement uncertainty; 1 kHz; 10 to 100 MHz; LED; RF attenuation measurement system; attenuation standard; dual-channel IF substitution method; intermediate frequency substitution method; light emitting diode; optical fiber; photodetector; reference standard; traceable inductive voltage divider; Assembly systems; Attenuation measurement; Frequency conversion; Light emitting diodes; Measurement standards; Optical attenuators; Optical frequency conversion; Standards development; Uncertainty; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2003.810010