The Measurement of Gamma Dose with Silicon Junction Counters
Author
Jones, A.R.
Author_Institution
Atomic Energy of Canada Limited, Biology and Health Physics Division, Chalk River, Ontario
Volume
9
Issue
5
fYear
1962
Firstpage
17
Lastpage
22
Abstract
Measurements have been made on three types of silicon Junction counters to assess their application in gamma ray dosimetry. These measurements are presented and it is concluded that each of the three types (high resistivity and guard-ring p-n junction counters and lithium drift counters) are suitable. Their limitations and means of application are also discussed.
Keywords
Atomic measurements; Counting circuits; Dosimetry; Electrons; Energy measurement; Lithium; P-n junctions; Pulse amplifiers; Radiation detectors; Silicon;