Title :
A Precision Digital Instantaneous Heart Rate Meter with a Wide Range
Author :
Yonezawa, Yoshiharu ; Ninomiya, Ishio ; Nishiura, Naoki ; Matsushita, Gohta
Author_Institution :
Department of Electrical Engineering, Hiroshima Institute of Technology
Abstract :
A precision digital instantaneous heart rate meter with a wide range of 0-1023 beats/min was developed for converting the time interval between two successive heart beats to frequency. The resolution of this device is 1 beat/min, and the maximum nonlinearity error is less than 0.06 percent relative to full scale. Therefore, this device has two advantages, i.e., it is more precise and it has a wider range than other conventional tachometers.
Keywords :
Animals; Biomedical measurements; Clocks; Counting circuits; Electrocardiography; Frequency conversion; Heart rate; Humans; Pulse generation; Voltage; Animals; Heart Function Tests; Heart Rate; Humans;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1983.325059