• DocumentCode
    1212435
  • Title

    A Hyperthermia System Using a New Type of Inductive Applicator

  • Author

    Anderson, J.Bach ; Baun, Aage ; Harmark, Kristian ; Heinzl, Leif ; Raskmark, Povl ; Overgaard, Jens

  • Author_Institution
    Institute of Electronic Systems, Aalborg University
  • Issue
    1
  • fYear
    1984
  • Firstpage
    21
  • Lastpage
    27
  • Abstract
    A hyperthermia system working at frequencies around 150 MHz is described. It is useful for a penetration depth of 3-4 cm in tumor and muscle tissue. The system includes a new type of inductive applicator giving a beam-like distribution in the tissue, a controllable surface cooling system, and multipoint temperature measurements. A phase I clinical study shows the efficacy of the system, but also the importance of temperature measurements at many points since, in general, the temperature gradients are large.
  • Keywords
    Applicators; Cancer; Control systems; Electromagnetic heating; Frequency; Hyperthermia; Muscles; Neoplasms; Temperature control; Temperature measurement; Carcinoma; Head and Neck Neoplasms; Humans; Hyperthermia, Induced; Models, Structural; Muscles; Neoplasms; Time Factors;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1984.325366
  • Filename
    4121710