DocumentCode
1212823
Title
AC-DC transfer comparator for the calibration of thermal voltage converters against Josephson alternating voltage standards
Author
Budovsky, Ilya ; Sasaki, Hitoshi ; Coogan, Peter
Author_Institution
Nat. Meas. Lab., CSIRO Telecommun. & Ind. Phys., Lindfield, NSW, Australia
Volume
52
Issue
2
fYear
2003
fDate
4/1/2003 12:00:00 AM
Firstpage
538
Lastpage
541
Abstract
This paper describes a new measurement system for the comparison of alternating voltage standards based on the Josephson effect with thermal voltage converters. The target is to achieve an uncertainty of 0.1 μV/V at frequencies up to 10 kHz. The present work achieves an uncertainty of 0.2 μV/V up to 2 kHz.
Keywords
Josephson effect; calibration; comparators (circuits); convertors; measurement standards; measurement uncertainty; voltage measurement; 2 kHz; AC-DC transfer comparator; Josephson alternating voltage standard; calibration; measurement uncertainty; thermal voltage converter; AC generators; Analog-digital conversion; Calibration; DC generators; Digital-analog conversion; Josephson effect; Measurement standards; Switches; Uncertainty; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2003.811671