DocumentCode :
1212828
Title :
The Whole Body 11.7 T MRI Magnet for Iseult/INUMAC Project
Author :
Vedrine, P. ; Aubert, G. ; Beaudet, F. ; Belorgey, J. ; Beltramelli, J. ; Berriaud, C. ; Bredy, P. ; Chesny, Ph ; Donati, A. ; Gilgrass, G. ; Grunblatt, G. ; Juster, F.P. ; Molinie, F. ; Meuris, C. ; Nunio, F. ; Payn, A. ; Quettier, L. ; Rey, J.M. ; Schil
Author_Institution :
DAPNIA, CEA/Saclay, Gif-sur-Yvette
Volume :
18
Issue :
2
fYear :
2008
fDate :
6/1/2008 12:00:00 AM
Firstpage :
868
Lastpage :
873
Abstract :
Neurospin, a neuroscience research centre with very high field MRI equipments, just opened in November 2006 at Saclay by the CEA life science division. One of the imaging systems will require an 11.7 T magnet with a 900 mm warm bore. This magnet is currently under development at CEA Saclay, in collaboration with Siemens Medical Solutions and Alstom Magnets and Superconductors, within the framework of the French-German consortium Iseult/INUMAC (Imaging of Neuro disease Using high field. MAgnetic resonance and Contrastophores). The main aim of the consortium is to promote magnetic resonance and molecular imaging within high magnetic fields. The proposed magnet design is based on conservative options, but definitely unusual construction methods, for an MRI magnet (pancake winding, liquefier, stabilized power supply). These key design points therefore need to be assessed with several prototypes, integrated within a 5 years projected development plan, ending in 2011. The paper will present the objectives of the project as well as the main characteristics of the magnet and its development plan.
Keywords :
biomedical MRI; biomedical equipment; diseases; molecular biophysics; neurophysiology; Alstom magnets; Alstom superconductors; CEA life science division; French-German consortium; Iseult/INUMAC project; MRI imaging systems; Neurospin; Saclay; Siemens Medical Solutions; high field MRI equipments; magnetic flux density 11.7 T; magnetic resonance-and-contrastophores; molecular imaging; neuro disease imaging; neuroscience research centre; pancake winding; size 900 mm; stabilized power supply; whole body MRI magnet; Magnetic resonance imaging; niobium titanium; superconducting magnet;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2008.920786
Filename :
4512185
Link To Document :
بازگشت