Abstract
We have used a quadrature bridge to compare directly two quantum Hall resistors with two 10-nF capacitors at 1233 Hz. This approach eliminates uncertainties due to thermal and structural instabilities and to AC/DC differences of the wire resistors normally used in an intermediate step. We have measured the capacitative interaction between the devices, and have developed a shielded header to suppress this effect. The value for the National Research Council primary standard 10-pF capacitor obtained by this method agrees with the expected value to within 0.1±3.5 aF, or (0.01±0.35)×10-6.
Keywords
Hall effect devices; bridge circuits; capacitance measurement; electromagnetic shielding; measurement standards; quantum Hall effect; resistors; thermal stability; 10 nF; 1233 Hz; AC bridges; National Research Council primary standard capacitor; capacitative interaction; capacitors; direct capacitance standards determination; quadrature bridge; quantized Hall resistor pair; shielded header; structural instabilities; thermal instabilities; Bridge circuits; Capacitance measurement; Capacitors; Councils; Laboratories; Measurement standards; Quantum capacitance; Resistors; Voltage; Wire;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2003.810454
Filename
1202097
Link To Document