• DocumentCode
    1212874
  • Title

    Direct determination of capacitance standards using a quadrature bridge and a pair of quantized Hall resistors

  • Author

    Inglis, A.D. ; Wood, B.M. ; Côté, M. ; Young, R.B. ; Early, M.D.

  • Author_Institution
    Inst. for Nat. Meas. Stand., Nat. Res. Council, Ottawa, Ont., Canada
  • Volume
    52
  • Issue
    2
  • fYear
    2003
  • fDate
    4/1/2003 12:00:00 AM
  • Firstpage
    559
  • Lastpage
    562
  • Abstract
    We have used a quadrature bridge to compare directly two quantum Hall resistors with two 10-nF capacitors at 1233 Hz. This approach eliminates uncertainties due to thermal and structural instabilities and to AC/DC differences of the wire resistors normally used in an intermediate step. We have measured the capacitative interaction between the devices, and have developed a shielded header to suppress this effect. The value for the National Research Council primary standard 10-pF capacitor obtained by this method agrees with the expected value to within 0.1±3.5 aF, or (0.01±0.35)×10-6.
  • Keywords
    Hall effect devices; bridge circuits; capacitance measurement; electromagnetic shielding; measurement standards; quantum Hall effect; resistors; thermal stability; 10 nF; 1233 Hz; AC bridges; National Research Council primary standard capacitor; capacitative interaction; capacitors; direct capacitance standards determination; quadrature bridge; quantized Hall resistor pair; shielded header; structural instabilities; thermal instabilities; Bridge circuits; Capacitance measurement; Capacitors; Councils; Laboratories; Measurement standards; Quantum capacitance; Resistors; Voltage; Wire;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2003.810454
  • Filename