DocumentCode :
1213169
Title :
System Optimization for Improved Accuracy in Transcutaneous Signal and Power Transmission
Author :
Hochmair, Erwin S.
Author_Institution :
Institut fuer Allgemneine Elektrotechnik und Elektronik, Technical University of Vienna
Issue :
2
fYear :
1984
Firstpage :
177
Lastpage :
186
Abstract :
A stable transcutaneous transmission of power and signal via two coupled coils with minimized dependence on the relative spacing of external and implanted coil is possible by employing "critical coupling" between implanted and external circuits. Optimizing coil geometries and preventing the RF output amplifier from saturating is necessary to approximately maintain critical coupling despite placement tolerances within a reasonable range. Based on these considerations a transcutaneous signal transmission system for an auditory prosthesis has been designed. This transmission system can also be used for several other stimulation purposes needing accurate control of stimulation parameters.
Keywords :
Absorption; Coils; Coupling circuits; Equivalent circuits; Geometry; Power supplies; Power transmission; Radio frequency; Radiofrequency amplifiers; Tuned circuits; Cochlear Implants; Electric Power Supplies; Electronics, Medical; Prosthesis Design;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1984.325404
Filename :
4121801
Link To Document :