Two methods of measuring ocular torsion from digital images of the eyes were developed and tested. One method measures torsion from the translation of two landmarks using a rectilinear coordinate system. The second method measures torsion from the translation of two landmarks using a polar coordinate system. The center for the polar sampling is the center of the pupil. After thresholding and filtering the images, landmark translation is measured from the interpolated peak in the normalized cross correlation of the reference landmark with the image.
Keywords :
Coordinate measuring machines; Cornea; Digital signal processing; Extraterrestrial measurements; Eyes; Image sampling; Irrigation; Lenses; Signal to noise ratio; Space exploration; Biomedical Engineering; Biometry; Electrooculography; Eye Movements; Humans;