DocumentCode :
1214250
Title :
Analysis of a Digital EMG Signal Processor in Dynamic Conditions
Author :
Alessio, T.D.
Author_Institution :
Dipartimento di Scienza e Tecnica dell´´lnformazione e della Comunicazione, Universiti di Roma
Issue :
1
fYear :
1985
Firstpage :
78
Lastpage :
82
Abstract :
The performances of digital electromyographic signal processors in dynamic conditions are determined by evaluating the root-mean-square value of the estimation error, including bias and variance of the estimates. The results thus obtained are valid for a wide class of myoprocessor schemes. The analysis has been used to optimize myoprocessors with moving-average smoothing filters, the length of which will be adapted to the local variability characteristics of the force in order to obtain a minimum overali estimation error.
Keywords :
Data acquisition; Electromyography; Fatigue; Frequency; Humans; Magnetic force microscopy; Monitoring; Signal analysis; Signal processing; White noise; Biomedical Engineering; Electromyography; Humans;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1985.325635
Filename :
4121932
Link To Document :
بازگشت