• DocumentCode
    1214276
  • Title

    New Heat Rate Meter Based on Frequency Grading

  • Author

    Basano, L. ; Ottonello, P.

  • Author_Institution
    Istituto di Scienze Fisiche, Universita di Genova
  • Issue
    1
  • fYear
    1985
  • Firstpage
    89
  • Lastpage
    91
  • Abstract
    We describe a wide-range (15-1000 bpm) and high-resolution (1 bpm) heart rate meter whose design is based on the automatic grading of the clock frequency. Each pulse resets the counter that starts counting at high frequency (16 667 Hz); if no pulse arrives within a preselected interval, the clock frequency is reduced by a factor of 4. This procedure may be repeated one or more times, depending on the pulse delay, and results in a step-like decrement of the clock frequency. This allows a very efficient use of the memory capabilities of two suitably programmed EPROM´s.
  • Keywords
    Animals; Clocks; Counting circuits; EPROM; Frequency; Heart rate; Human resource management; Intellectual property; Physiology; Pulse measurements; Animals; Biomedical Engineering; Heart Rate; Humans;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1985.325638
  • Filename
    4121935