Title
Design and Evaluation of a Digital Closed-Loop Controller for the Regulation of Muscle Force by Recruitment Modulation
Author
Wilhere, George F. ; Crago, Patrick E. ; Chizeck, Howard J.
Author_Institution
Applied Neural Control Laboratory, Departments of Biomedical Engineering and Systems Engineering, Case Western Reserve University
Issue
9
fYear
1985
Firstpage
668
Lastpage
676
Abstract
In principle, orthoses employing electrically stimulated muscles can be improved by the incorporation of closed-loop control systems to regulate muscle output. One obstacle to attaining acceptable closed-loop control systems is the muscle recruitment characteristic, a memoryless, nonlinear, time-varying relationship between the recruitment modulation stimulus parameter (pulsewidth or amplitude) and the muscle force.
Keywords
Amplitude modulation; Control systems; Digital control; Digital modulation; Force control; Muscles; Nonlinear control systems; Orthotics; Recruitment; Time varying systems; Animals; Biomedical Engineering; Cats; Electric Stimulation; Humans; Models, Biological; Muscle Contraction; Orthotic Devices;