DocumentCode :
1216192
Title :
Neurointensive care unit system for continuous electrophysiological monitoring with remote web-based review
Author :
Van der Kouwe, André J W ; Burgess, Richard C.
Author_Institution :
Cleveland Clinic Found., OH, USA
Volume :
7
Issue :
2
fYear :
2003
fDate :
6/1/2003 12:00:00 AM
Firstpage :
130
Lastpage :
140
Abstract :
There is a need in the neurological intensive care unit for a single integrated bedside monitor for continuously monitoring the function of the patient´s central nervous system. In this paper, we demonstrate the feasibility of building such a system and operating it in the intensive care environment. We have developed a fully automated system that samples electrophysiological waveforms of various modalities according to a schedule of predefined intervals along with routinely monitored cardiac and respiratory parameters. The system provides stimulation and acquires responses without requiring supervision. The electrophysiological data include brainstem auditory and somatosensory evoked potentials and epochs of the electroencephalogram. The system applies peak detection and spectral analysis to extract salient parameters from the raw waveforms. The results are made available immediately in real time on the local network for local review and further analysis. A Web-based interface makes review by a qualified neurologist possible anywhere within the hospital´s secure intranet during and after monitoring. This system could potentially give an early warning of impending herniation, subclinical seizures, and brain or spinal cord ischemia. We demonstrate its application in a few diverse neurological intensive care cases and a case in the interventional neuroradiology suite.
Keywords :
auditory evoked potentials; computerised monitoring; electroencephalography; medical computing; somatosensory phenomena; brain ischemia; brainstem auditory potentials; central nervous system; electroencephalogram; electrophysiological waveforms; impending herniation; intensive care environment; interventional neuroradiology suite; neurological intensive care cases; neurological intensive care unit; routinely monitored cardiac parameters; routinely monitored respiratory parameters; secure intranet; single integrated bedside monitor; somatosensory evoked potentials; spinal cord ischemia; subclinical seizures; Central nervous system; Computerized monitoring; Condition monitoring; Data mining; Electrophysiology; Ischemic pain; Patient monitoring; Remote monitoring; Spectral analysis; Spinal cord; Aged; Decision Making, Computer-Assisted; Diagnosis, Computer-Assisted; Diagnostic Techniques, Neurological; Electroencephalography; Evoked Potentials; Female; Humans; Information Storage and Retrieval; Intensive Care; Internet; Monitoring, Physiologic; Ohio; Pilot Projects; Point-of-Care Systems; Systems Integration; Telemedicine; User-Computer Interface;
fLanguage :
English
Journal_Title :
Information Technology in Biomedicine, IEEE Transactions on
Publisher :
ieee
ISSN :
1089-7771
Type :
jour
DOI :
10.1109/TITB.2003.811873
Filename :
1203142
Link To Document :