DocumentCode :
1216692
Title :
Performance of a BaF/sub 2/-TMAE prototype detector for use in PET
Author :
Wells, K. ; Visvikis, D. ; Ott, R.J. ; Zweit, J. ; Bateman, J.E. ; Stephenson, R. ; Connolly, J. ; Tappern, G.
Author_Institution :
Dept. of Phys., R. Marsden Hospital, Sutton, UK
Volume :
41
Issue :
6
fYear :
1994
Firstpage :
2737
Lastpage :
2742
Abstract :
A BaF/sub 2/-TMAE detector has been developed for use in positron emission tomography (PET). The detector achieves an efficiency of 23.5%, a time resolution of 3.5 ns FWHM, and a spatial resolution of 6.0 mm FWHM with no change in the normal operating parameters. The only deterioration observed in a typical working day is a small increase in the FWTM of the time resolution. This paper describes how this level of performance was achieved and also how these results have been used as the basis for the design of a large area positron camera.<>
Keywords :
positron emission tomography; solid scintillation detectors; 3.5 ns; 6 mm; BaF/sub 2/; BaF/sub 2/-TMAE detector; FWHM; FWTM; PET; efficiency; large area positron camera; positron emission tomography; spatial resolution; time resolution; Computational fluid dynamics; Delay lines; Electromagnetic wave absorption; Electrons; Positron emission tomography; Prototypes; Pulse amplifiers; Scintillation counters; Solid scintillation detectors; Spatial resolution;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.340641
Filename :
340641
Link To Document :
بازگشت