DocumentCode :
1216766
Title :
Investigation of the use of X-ray CT images for attenuation compensation in SPECT
Author :
LaCroix, K.J. ; Tsui, B.M.W. ; Hasegawa, B.H. ; Brown, J.K.
Author_Institution :
Dept. of Biomed. Eng., North Carolina Univ., Chapel Hill, NC, USA
Volume :
41
Issue :
6
fYear :
1994
Firstpage :
2793
Lastpage :
2799
Abstract :
This study investigates the general use of single-beam X-ray computed tomography (CT) images for generating attenuation maps for compensation of photon attenuation in SPECT images. A 3D mathematical thorax phantom is used to simulate both emission and transmission projection data for monoenergetic (radionuclide) and polyenergetic (X-ray) sources. Polyenergetic transmission projection data are simulated for a standard X-ray spectrum and fan-beam geometry. The projection data are reconstructed using filtered backprojection to form an X-ray CT image which is then scaled to produce an estimate of the attenuation map at the energy of the emission radionuclide. Emission projection data are simulated for a fan-beam geometry at the energies of /sup 201/Tl and /sup 99m/Tc, two radionuclides commonly used in cardiac SPECT. Detector response and scatter are not included in the model. Noiseless, emission projection data are iteratively reconstructed using the ML-EM algorithm with nonuniform attenuation compensation and attenuation maps derived from both the simulated X-ray CT image and from a simulated monoenergetic transmission CT image. The attenuation maps generated from the X-ray CT images accurately estimate the attenuation coefficient for muscle and lung tissues, but not for bone tissues, which show error in the attenuation coefficient of 21-42% for spinal bone and 34-58% for rib bone. However, despite the inaccurate estimate of bone attenuation, the reconstructed SPECT images provide estimates of myocardial radioactivity concentration to within 9% and show few artifacts.<>
Keywords :
computerised tomography; medical image processing; single photon emission computed tomography; /sup 201/Tl; /sup 99m/Tc; 3D mathematical thorax phantom; ML-EM algorithm; SPECT attenuation compensation; Tc; Tl; X-ray CT images; attenuation maps; bone; fan-beam geometry; lung; monoenergetic source; muscle; polyenergetic source; single-beam X-ray computed tomography images; standard X-ray spectrum; Attenuation; Bones; Computational modeling; Computed tomography; Geometry; Image generation; Image reconstruction; Single photon emission computed tomography; Solid modeling; X-ray imaging;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.340649
Filename :
340649
Link To Document :