• DocumentCode
    1217556
  • Title

    Real-Time Fetal ECG Recording

  • Author

    Bergveld, Piet ; KÖlling, Arjan J. ; Peuscher, Jan H.J.

  • Author_Institution
    Biomedical Engineering Division, Department of Electrical Engineering, Twente University of Technology
  • Issue
    5
  • fYear
    1986
  • fDate
    5/1/1986 12:00:00 AM
  • Firstpage
    505
  • Lastpage
    509
  • Abstract
    We present a real-time abdominal recording method for the FECG based on a minimization procedure for the MECG interference, by weighted addition of four abdominal leads containing FECG with a large amount of MECG. The minimization of the MECG does not influence the FECG recording adversely. FECG complexes coinciding with MECG complexes are also recorded. Use is made of programmable amplifiers that multiply analog abdominal leads by digital coefficients. These coefficients are calculated at the beginning of the recording by a microcomputer system. Experimental results are shown that clarify the advantages of the measurement principle. The real-time recorded FECG is wide-banded, so that all generally known ECG signal processing techniques, such as complex averaging, can be performed.
  • Keywords
    Abdomen; Biomedical measurements; Electrocardiography; Electrodes; Fetal heart; Fetal heart rate; Minimization methods; Position measurement; Ultrasonic imaging; Ultrasonic variables measurement; Electrocardiography; Female; Fetal Heart; Fetal Monitoring; Humans; Pregnancy;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1986.325737
  • Filename
    4122324