• DocumentCode
    1217947
  • Title

    Dynamic response of a frequency measuring system

  • Author

    Prokin, Milan

  • Author_Institution
    Fac. of Electr. Eng., Belgrade Univ., Yugoslavia
  • Volume
    41
  • Issue
    3
  • fYear
    1992
  • fDate
    6/1/1992 12:00:00 AM
  • Firstpage
    390
  • Lastpage
    396
  • Abstract
    A frequency measurement system using a double-buffered method is designed for PC/AT compatible computers. The extremely wide range of measured frequencies (from 1.67 mHz to 7 MHz) is independent of a wide range of sampling periods (from 100 μs to 300 s). The response time to an instantaneously increased frequency from 1.67 mHz to 7 MHz is less than two sampling periods. The menu-driven software graphically presents frequency versus time to analyze fast frequency change using rapid sampling, or to measure the average frequency during long sampling periods
  • Keywords
    IBM compatible machines; computerised instrumentation; dynamic response; frequency measurement; graphical user interfaces; 1.67 mHz to 7 MHz; 100 mus to 300 s; PC/AT compatible computers; average frequency; double buffered technique; double-buffered method; dynamic response; frequency change; menu-driven software; sampling; Clocks; Counting circuits; Delay; Design methodology; Frequency measurement; Frequency synchronization; Pulse circuits; Pulse measurements; Sampling methods; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.153335
  • Filename
    153335