• DocumentCode
    1218168
  • Title

    Conversion Between Plethysmograph and Perturbational Airways Resistance Measurements

  • Author

    Johnson, Arthur T.

  • Author_Institution
    Department of Agricultural Engineering, Unversity of Maryland
  • Issue
    8
  • fYear
    1986
  • Firstpage
    803
  • Lastpage
    806
  • Abstract
    Equations were developed for airway pressure-flow characteristics which were then used to illustrate differences between airflow perturbation device measurements and body plethysmographic airways resistance measurements. The general conclusions indicate that perturbational resistances may be able to be reformulated as plethysmographic resistances, yielding information where the body plethysmograph itself is not well suited.
  • Keywords
    Agricultural engineering; Biomedical measurements; Difference equations; Electrical resistance measurement; Fluid flow measurement; Force measurement; Immune system; Lungs; Volume measurement; Airway Resistance; Biomedical Engineering; Humans; Plethysmography, Whole Body;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1986.325910
  • Filename
    4122396