DocumentCode :
1219540
Title :
Enhancement and Dynamic Time Warping of Somatosensory Evoked Potential Components Applied to Patients with Multiple Sclerosis
Author :
Roberts, Kim ; Lawrence, Peter ; Eisen, Andrew ; Hoirch, Maureen
Author_Institution :
Department of Electrical Engineering, University of British Columbia
Issue :
6
fYear :
1987
fDate :
6/1/1987 12:00:00 AM
Firstpage :
397
Lastpage :
405
Abstract :
A zero-phase, high-pass digital filter at 300 Hz was used to emphasize the higher frequency, smaller peaks of the somatosensory evoked potential (SEP). Using the digital filter, a consistent pattern of peaks was observed in normals. A dynamic time-warping technique was created to align the peaks of digitally high-pass filtered SEP´s. Based on this technique, a clinical test for abnormality of the SEP was developed and applied to 25 multiple sclerosis patients. The results showed that three times as many multiple sclerosis SEP´s were identified by the dynamic time warping as by the traditional latency measurement.
Keywords :
Delay; Digital filters; Electrical stimulation; Filtering; Frequency; Helium; Multiple sclerosis; Optical fiber sensors; Testing; Time measurement; Adolescent; Adult; Evoked Potentials; Female; Humans; Male; Multiple Sclerosis; Somatosensory Cortex;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1987.326055
Filename :
4122561
Link To Document :
بازگشت