DocumentCode :
1219625
Title :
Cardiac Mapping Instrumentation for the Instantaneous Display of Endocardial and Epicardial Activation
Author :
Parson, Ian D. ; Downar, Eugene
Author_Institution :
Institute of Biomedical Engineering and the Department of Medicine, University of Toronto
Issue :
6
fYear :
1987
fDate :
6/1/1987 12:00:00 AM
Firstpage :
468
Lastpage :
472
Abstract :
A real-time mapping system is described that allows instant display of both endocardial and epicardial activation as monitored by an array of transmural plunge-needle electrodes in the intact in situ heart. From the total cardiac muscle mass, 768 electrograms are recorded, with a programmable feature allowing the rapid and unrestricted selection of 110 endocardial electrograms for automated analysis. An activation analyzer (consisting of parallel analog circuitry) provides an activation image in real time, while video recorders provide both raw and processed data storage with the capability of immediate slow-motion replay. The nature of the data acquisition (110 plunge needles) limits this system to the experimental arrhythmia laboratory where its chief advantage is the enabling of the multichannel recording of endo data without the complications of open heart surgery. Equally important is the ability to automatically store the endocardial activation image locked in time to the simultaneously recorded epicardial image.
Keywords :
Circuits; Data acquisition; Displays; Electrodes; Heart; Image analysis; Instruments; Memory; Muscles; Real time systems; Arrhythmias, Cardiac; Computers; Electrocardiography; Endocardium; Heart; Humans;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.1987.326082
Filename :
4122570
Link To Document :
بازگشت