Title :
Optimal System Design for an Implantable CW Doppler Ultrasonic Flowmeter
Author :
Di Pietro, David M. ; Meindl, James D.
Author_Institution :
Hewlett-Packard Company, Santa Clara Division
fDate :
5/1/1978 12:00:00 AM
Abstract :
A totally implantable telemetry system utilizing custom micropower monolithic integrated circuits has been developed to measure instantaneous pulsatile blood flow in the major arteries of animals. The device utilizes the Doppler effect at ultrasonic frequencies to achieve a flowmeter possessing small size and weight, low power consumption, inherent accuracy, and long-term stability. The integrated circuit electronics package occupies a volume of 3.8 cc and requires a power consumption of 10 mW from a single 135 volt mercury cell. The volume of the entire implantable flowmeter package is less than 36 cc. As a micro-miniature radio-telemetry system, the flow-meter can be totally implanted within the body, thereby eliminating the need for wires penetrating the skin, a serious problem with "back-pack" flowmeters due to danger of infection. The implantable flow-meter system provides a new and necessary tool for accurate quantitative measurements in the study of cardiovascular disease. It has been used in the study of transplanted heart rejection to monitor blood velocity in the major arteries of doB
Keywords :
Arteries; Blood flow; Electronics packaging; Energy consumption; Fluid flow measurement; Integrated circuit measurements; Integrated circuit packaging; Monolithic integrated circuits; Telemetry; Ultrasonic variables measurement; Animals; Aorta; Blood Flow Velocity; Cardiovascular Diseases; Dogs; Doppler Effect; Models, Biological; Radio; Rheology; Telemetry; Ultrasonography;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1978.326330