Abstract :
This communication describes results of preliminary testing of a new method of implementing EEG waveform detection. The circuit described is capable of detecting EEG alpha, beta, delta, sleep spindles, muscle artifact, etc., on three separate channels simultaneously when operating at 32: 1 time compression. Microprocessor implementation allows for adaptive control of detection parameters, making hardware changes unnecessary.