• DocumentCode
    1221947
  • Title

    Slew Rates for Electrocardiographic Signals

  • Author

    Berson, Alan S. ; Pipberger, Hubert V.

  • Author_Institution
    Veterans Administration Hospital, and the Departments of Clinical Engineering and Medicine, The George Washington University
  • Issue
    3
  • fYear
    1978
  • fDate
    5/1/1978 12:00:00 AM
  • Firstpage
    299
  • Lastpage
    301
  • Abstract
    Frank lead electrocardiograms from 100 adults and 61 infants were recorded on magnetic tape and digitized at 2 ms intervals after bandwidth limiting fmst at 200 Hz and then at 100 Hz. Slew rates based on voltage differences between successive sampled values during QRS were computed for one cardiac cycle for each lead. Considered as a group, 200 ¿V/ms rates occur about 1% of the time in adult records, and 400 ¿V/ms rates occur about 1% of the time in infants´ records. In extreme individual records, 300 ¿V/ms rates were exceeded 2% of the time in an adult record, and 400 ¿V/ms rates were exceeded 6.5% of the time in an infant´s record. No significant differences in slew rates were observed to be caused by variation in bandwidth from 100 Hz to 200 Hz. It is recommended that ECG instruments be designed to be capable of faithfully reproducing at least 400 ¿V/ms slew rates.
  • Keywords
    Bandwidth; Electrocardiography; Frequency response; Heart; Instruments; Low pass filters; Lungs; Pediatrics; Sampling methods; Voltage; Adult; Electricity; Electrocardiography; Female; Humans; Infant; Male; Time Factors;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1978.326347
  • Filename
    4122832