• DocumentCode
    1221956
  • Title

    A Fluidic Drive System for Intra-Aortic Balloon Heart Assist Device

  • Author

    Hariharan, R. ; Menon, K.A.P.

  • Author_Institution
    Department of Mechanical Engineering, Indian Institute of Science
  • Issue
    3
  • fYear
    1978
  • fDate
    5/1/1978 12:00:00 AM
  • Firstpage
    301
  • Lastpage
    303
  • Abstract
    Intra-aortic balloon pumping is a counter pulsation technique for temporary circulatory assistance in cardiogenic shock and other low cardiac output conditions. Conventional systems use a balloon at the end of a catheter driven by a solenoid valve, controlled by patient´s ECG or ventricular pressure signal. This results in time delay introducted by solenoid spool inertia, gas inertia, and hysteresis effects of the solenoid. Fluidics, because of their non-moving part operation and high switching speeds, minimizes the inertial effects while contributing high reliability. This communication describes a fluidic system developed for driving the balloon accepting electric control signals.
  • Keywords
    Cardiology; Catheters; Communication system control; Control systems; Counting circuits; Electric shock; Heart; Pumps; Solenoids; Valves; Assisted Circulation; Intra-Aortic Balloon Pumping; Shock, Cardiogenic;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.1978.326348
  • Filename
    4122833