DocumentCode :
1222718
Title :
An Improved Correction of Distorted Indicator-Dilution Curves
Author :
Nassi, M. ; Dayan, J. ; Braun, S. ; Weinstein, H.
Author_Institution :
Department of Electrical Engineering, Stanford University
Issue :
1
fYear :
1979
Firstpage :
18
Lastpage :
21
Abstract :
Venoarterial indicator dilution curves are distorted by non-ideal injection and by the monitoring system. An improved method of correcting these curves is developed assuming that a linear model of perfect mixers in series can be fitted to the physiologic system response as well as to the subsystem responses.
Keywords :
Biomedical monitoring; Chemical technology; Dispersion; Distortion measurement; Error correction; Helium; Mechanical engineering; Sampling methods; Shape; Time measurement; Indicator Dilution Techniques; Models, Biological; Monitoring, Physiologic;
fLanguage :
English