Construction and operation of a low temperature bolometer for detection of gamma-rays
Author :
Alessandrello, A. ; Brofferio, C. ; Camin, D.V. ; Fiorini, E. ; Giuliani, A.
Author_Institution :
Dipartimento di Fisica, Istituto Nazionale di Fisica Nucl., Milan, Italy
Volume :
36
Issue :
1
fYear :
1989
fDate :
2/1/1989 12:00:00 AM
Firstpage :
141
Lastpage :
145
Abstract :
The authors report the first measurement of high-energy gamma rays with a low-temperature calorimeter. The detector is a crystal of about 10.7 gm operated at a temperature of about 30 mK. The thermal pulses produced by gamma rays are detected and measured by a thermistor in thermal contact with the crystal. Spectra obtained by exposure to gamma rays of energies ranging from 122 to 2615 keV are reported and discussed
Keywords :
bolometers; gamma-ray detection and measurement; particle detectors; 122 to 2615 keV; 30 mK; calorimeter; high-energy gamma rays; low temperature bolometer; thermal pulses; thermistor; Bolometers; Energy resolution; Gamma ray detection; Gamma ray detectors; Phonons; Physics; Pulse measurements; Semiconductor diodes; Temperature; Thermistors;