DocumentCode
1223317
Title
Single Frequency RF Powered ECG Telemetry System
Author
Ko, Wen H. ; Hyneck, Jaroslav ; Homa, John
Author_Institution
Engineering Design Center, Case Western Reserve University
Issue
2
fYear
1979
Firstpage
105
Lastpage
109
Abstract
It has been demonstrated that a radio frequency magnetic field can be used to power implanted electronic circuitry for short range telemetry to replace batteries. A substantial reduction in implanted volume can be achieved by using only one RF tank circuit for receiving the RF power and transmitting the telemetered information. A single channel telemetry system of this type, using time sharing techniques, was developed and employed to transmit the ECG signal from Rhesus monkeys in primate chairs. The signal from the implant is received during the period when the RF powering radiation is interrupted. The ECG signal is carried by 20 , us pulse position modulated pulses, referred to the trailing edge of the RF powering pulse. satisfactory results have been obtained with this single frequency system. The concept and the design presented may be useful for short range, longterm implant telemetry systems.
Keywords
Batteries; Electrocardiography; Implants; Magnetic circuits; Magnetic fields; Pulse modulation; RF signals; Radio frequency; Telemetry; Time sharing computer systems; Amplifiers; Animals; Electrocardiography; Electronics, Medical; Haplorhini; Macaca mulatta; Radio Waves; Telemetry;