Title :
A precision pulse generator for TDC calibration
Author :
Sasaki, O. ; Taniguchi, T. ; Ikeda, M. ; Ohska, T.K. ; Miyazawa, M.
Author_Institution :
Nat. Lab. for High Energy Phys., Ibaraki, Japan
fDate :
2/1/1989 12:00:00 AM
Abstract :
A CAMAC high-accuracy calibration source module for high-precision TDCs (time-to-digital converters) was developed. It was found that the module has a 14-bit accuracy (linearity) of the time interval between START and STOP output pulses (i.e. as low as 13 ps at 200 ns full range) with less than 15-ps jitter. Full range is switch-selectable (200 ns, 500 ns, 1000 ns and 2000 ns), and has 16-bit dynamic range
Keywords :
CAMAC; analogue-digital conversion; calibration; nuclear electronics; pulse generators; 14-bit accuracy; 16-bit dynamic range; CAMAC high-accuracy calibration source module; high-precision TDCs; jitter; output pulses; precision pulse generator; time-to-digital converters; CAMAC; Calibration; Circuit testing; Dynamic range; Jitter; Linearity; Pulse generation; Switches; Timing; Tuned circuits;
Journal_Title :
Nuclear Science, IEEE Transactions on