    Computation of Respiratory Resistance, Compliance, and Inertance from Forced Oscillatory Impedance Data

  • Author

    Tsai, Ming-Jer ; Pimmel, Russell L.

  • Author_Institution
    Massachusetts Institute of Technology, Lincoln Laboratory
  • Issue
    8
  • fYear
    1979
  • Firstpage
    492
  • Lastpage
    493
  • Abstract
    Relationships for computing total respiratory resistance, compliance, and inertance from the frequency dependence of forced oscillatory impedance are presented. These equations result by minimizing the sum squared difference between the experimental impedance data and the response of the model.
  • Keywords
    Biomedical measurements; Dogs; Equations; Force measurement; Frequency dependence; Frequency measurement; Immune system; Impedance measurement; Laboratories; Mathematical model; Airway Resistance; Humans; Lung Compliance; Models, Biological; Oscillometry; Respiratory Physiology;