A Simple Test for Phase Characteristics of Chart Recorders
Author :
Wagner, James W.
Author_Institution :
Research and Testing Staff, Bureau of Medical Devices, Food and Drug Administration, HEW/FDA/BMD Laboratory
Issue :
9
fYear :
1979
Firstpage :
505
Lastpage :
508
Abstract :
A method for quantifying the relative phase shift characteristics of chart recorders is presented. A test signal of the form (cos wot + A cos 3wot) is used, which when normalized for unequal responses at the two frequencies, provides an independent quantitative measure of relative phase shift which can be detennined from examination of the output record. Ihe ease of implementation and interpretation suggests the potential application of this test as both a compliance and quality assurance technique for evaluating the performance of medical chart recording systems.
Keywords :
Biomedical monitoring; Blood pressure; Frequency; Instruments; Medical diagnostic imaging; Medical tests; Passband; Phase distortion; Phase measurement; Testing; Electrodiagnosis; Electronics, Medical; Monitoring, Physiologic;