Defibrillation Recovery Curves for Different Electrode Materials
Author
Das, Dhruba P. ; Webster, John G.
Author_Institution
Department of Electrical and Computer Engineering, University of Wisconsin
Issue
4
fYear
1980
fDate
4/1/1980 12:00:00 AM
Firstpage
230
Lastpage
233
Abstract
We tested 16 electrode materials for their electrical characteristics and normalized all results to a 1 cm2 area. The electrode pair impedance versus frequency from 5 to 2000 Hz ranged from 20 to 58 000 ¿. We passed 2 mC through them in less than 5 ms to simulate defibrillation discharge. The resulting recovery curves showed voltage offsets after 5 s of from 300 to 800 mV.
Keywords
Biomedical electrodes; Defibrillation; Electrocardiography; Immune system; Impedance; Materials testing; Monitoring; Polarization; Tin; Voltage; Electric Countershock; Electrodes;