    Defibrillation Recovery Curves for Different Electrode Materials

  • Author

    Das, Dhruba P. ; Webster, John G.

  • Author_Institution
    Department of Electrical and Computer Engineering, University of Wisconsin
  • Issue
    4
  • fYear
    1980
  • fDate
    4/1/1980 12:00:00 AM
  • Firstpage
    230
  • Lastpage
    233
  • Abstract
    We tested 16 electrode materials for their electrical characteristics and normalized all results to a 1 cm2 area. The electrode pair impedance versus frequency from 5 to 2000 Hz ranged from 20 to 58 000 ¿. We passed 2 mC through them in less than 5 ms to simulate defibrillation discharge. The resulting recovery curves showed voltage offsets after 5 s of from 300 to 800 mV.
  • Keywords
    Biomedical electrodes; Defibrillation; Electrocardiography; Immune system; Impedance; Materials testing; Monitoring; Polarization; Tin; Voltage; Electric Countershock; Electrodes;