Title :
Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate Determination
Author :
Azevedo, Stephen ; Longini, Richard L.
Author_Institution :
Biomedical Engineering Program, Carnegie-Mellon University
fDate :
5/1/1980 12:00:00 AM
Abstract :
The reliability of fetal heart rate (FHR) from external abdominal-lead electrocardiographic waveforms depends upon the R-wave detection capability of the monitoring device. Assuming that the maternal component of the ECG is canceiled by techniques described elsewhere [1]-[4], we are left with a fetal electrocardiogm (FECG) which may be conrupted by high noise levels. This paper describes several digital signal processing techniques by which the detection of fetal R-waves is improved. These techniques-narrow band-pass filtering, local peak differencing, autoregressive filtering, and matched filtering-are theoretically developed. The performance of each technique is measured and analyzed for two sample fetal ECG inputs.
Keywords :
Abdomen; Band pass filters; Electrocardiography; Fetal heart rate; Filtering; Heart rate; Heart rate detection; Heart rate measurement; Matched filters; Noise level; Analog-Digital Conversion; Electrocardiography; Fetal Heart; Fetal Monitoring; Forecasting; Heart Rate; Humans; Image Enhancement;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.1980.326631