DocumentCode
1225607
Title
Design and evaluation of HEADTOME-IV, a whole-body positron emission tomograph
Author
Iida, Hidehiro ; Miura, Shuichi ; Kanno, Iwao ; Murakami, Matsutaro ; Takahashi, Kazuhiro ; Uemura, Kazuo ; Hirose, Yoshiharu ; Amano, Masaharu ; Yamamoto, Seiichi ; Tanaka, Kazumi
Author_Institution
Res. Inst. for Brain & Blood Vessels, Akita, Japan
Volume
36
Issue
1
fYear
1989
fDate
2/1/1989 12:00:00 AM
Firstpage
1006
Lastpage
1010
Abstract
A whole-body positron emission tomograph called Headtome-IV has been developed, and its physical performance has been investigated. An in-plane spatial resolution of 4.5 mm was realized with stationary sampling at the center of the field of view. The axial slice thicknesses were 9.5 and 9.0 mm for direct and cross planes, respectively. By moving the gantry framework axially, transaxial images of 14 or 21 slices are obtained quasi-simultaneously. The real-time large-scale cache memory system allowed real-time corrections for deadtime and radionuclide decay and real-time weighted integration for the purpose of a rapid calculation of rate-constant images
Keywords
biomedical equipment; computerised tomography; radioisotope scanning and imaging; HEADTOME-IV; deadtime; in-plane spatial resolution; radionuclide decay; rate-constant images; real-time corrections; real-time large-scale cache memory system; real-time weighted integration; stationary sampling; whole-body positron emission tomograph; Apertures; Biomedical imaging; Blood vessels; Cities and towns; Detectors; Image sampling; Large scale integration; Motion detection; Radioactive decay; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.34594
Filename
34594
Link To Document