DocumentCode
1225739
Title
Application of the Electrical Sizing Principle of Coulter to a New Multiparameter System
Author
Groves, Michael R.
Author_Institution
Coulter Electronics, Inc.
Issue
7
fYear
1980
fDate
7/1/1980 12:00:00 AM
Firstpage
364
Lastpage
369
Abstract
A new technique is described for simultaneously measuring, within 100 ¿s, cell volume, membrane breakdown voltage, and internal conductivity on individual cells of microscopic dimensions. The system is a modification of the hydrodynamically focusing Coulter Counter®. Tumor cells established from rat fetal brain cells of average volume 2100 ¿m3 were measured. A breakdown voltage of 0.86 V was found with an apparent volume underestimation after breakdown of 40 percent.