DocumentCode :
1225810
Title :
Axial sampling requirement for three-dimensional quantification of myocardial function with positron emission tomography
Author :
Raylman, R.R. ; Hutchins, G.D. ; Schwaiger, M. ; Paradise, A.H.
Author_Institution :
Div. of Nucl. Med., Michigan Univ. Med. Center, Ann Arbor, MI, USA
Volume :
36
Issue :
1
fYear :
1989
fDate :
2/1/1989 12:00:00 AM
Firstpage :
1030
Lastpage :
1033
Abstract :
The axial sampling requirements in cardiac positron emission tomography (PET) studies for the detection of myocardial defects has been investigated using stationary and beating heart phantoms. This study examines the relationship between defect volume quantification and axial sampling frequency in a stationary myocardium. The effect that myocardial motion has on this relationship is determined. The variation in volume and activity quantification as a function of defect position in the tomograph is discussed
Keywords :
cardiology; computerised tomography; radioisotope scanning and imaging; PET; axial sampling frequency; axial sampling requirements; beating heart phantoms; cardiac positron emission tomography; defect volume quantification; myocardial defects; stationary myocardium; Attenuation measurement; Contracts; Electromagnetic scattering; Heart; Imaging phantoms; Myocardium; Particle scattering; Positron emission tomography; Sampling methods; Shape measurement;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.34599
Filename :
34599
Link To Document :
بازگشت