Quality control in PET systems employing 2-D modular detectors
Author :
Daghighian, F. ; Hoffman, E.J. ; Huang, S.C.
Author_Institution :
Div. of Nucl. Med. & Biophys., California Univ., Los Angeles, CA, USA
Volume :
36
Issue :
1
fYear :
1989
fDate :
2/1/1989 12:00:00 AM
Firstpage :
1034
Lastpage :
1037
Abstract :
The authors have developed and put into operation three automated quality control procedures for PET (positron emission tomography) with 2-D modular detectors that are designed to detect problems quickly with a minimum effort on the part of the user. These procedures check the following: (1) the stability of the system in terms of the efficiencies of the detectors in a module or block; (2) resolution and its uniformity throughout the field of view; and (3) the stability of the whole system in terms of the data reproducibility. The stability of the detector modules is checked in terms of the uniformity of detector efficiencies by conducting a chi-square type test on the count rates of each detector in a block. In order to check the resolution and its uniformity and also the placement of the data in the PET scans, a cylindrical lucite phantom was fabricated which has a set of axial holes placed at 2-cm intervals. A computer program has been developed to analyze the images produced by this phantom and determine deviations from the normal resolution and correct placement of the data. A third procedure examines the reproducibility of the system by scanning a cylinder filled with a solution of Ge-68 several times and comparing the count rates in a small region of interest
Keywords :
automatic testing; calibration; computerised tomography; quality control; radioisotope scanning and imaging; 2-D modular detectors; 68Ge; PET; automated quality control procedures; chi-square type test; computer program; count rates; cylindrical lucite phantom; data reproducibility; positron emission tomography; resolution; uniformity; Detectors; Image resolution; Nuclear medicine; Photomultipliers; Positron emission tomography; Quality control; Reproducibility of results; Spatial resolution; Stability; Testing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.34600
Filename :
34600
Link To Document :
بازگشت