Abstract :
A method used to provide gain variation with received echo range to compensate for tissue attenuation in standard B-mode ultrasound imaging systems achieves signal multiplication by first logarithmically compressing the complete dynamic range of the received echo. The time-gain correction method described here is a modification of this method. It performs the time-gain correction operation digitally between the digital scan converter and the display. The overall system transfer function can be specified by means of a software interface without changing the characteristics of the analog front end. This interface can be automated to almost any degree, and it can include preset settings for typical imaging requirements. This method also allows the gain function to be changed even when new images are not being generated
Keywords :
acoustic imaging; biomedical ultrasonics; patient diagnosis; digital scan converter; logarithmic compression; medical diagnostic imaging; signal multiplication; software interface; standard B-mode ultrasound imaging; time-gain correction method; time-gain correction operation; tissue attenuation; Analog-digital conversion; Attenuation; Displays; Dynamic range; Envelope detectors; Gain; Image storage; Probes; Transfer functions; Ultrasonic imaging;
Journal_Title :