Title
Speckle in ultrasound B-mode scans
Author
Burckhardt, Christoph B.
Author_Institution
Central Res. Units, F Hoffmann-LaRoche and Co., Ltd., Basel, Switzerland
Volume
25
Issue
1
fYear
1978
Firstpage
1
Lastpage
6
Abstract
Ultrasound images obtained with a simple linear or sector scan show a granular appearance, called "speckle." This speckle is analyzed. The reduction in speckle that can be obtained with a compound scan with maximum amplitude writing is computed. The reduction in speckle is almost as large as can be obtained with averaging. It depends on the number of independent amplitude values that are measured. The condition for the independence of two amplitude values is derived, and thus a limit is given for the possible reduction in speckle.