• DocumentCode
    1227537
  • Title

    Multiple crossed-beam ultrasound Doppler velocimetry

  • Author

    Fox, Martin D.

  • Volume
    25
  • Issue
    5
  • fYear
    1978
  • Firstpage
    281
  • Lastpage
    286
  • Abstract
    A new ultrasound Doppler technique for obtaining calibrated 3D flow velocity information is reported. Theoretical expressions are derived for the expected Doppla shift and the spectral spread under a variety of flow conditions. Experimental results using a revolv-) ing turntable demonstrated the linearity of velocity detection, the ability to resolve an arbitrary velocity vector, and conrimed the theoretical predictions of spectral spread.
  • Keywords
    Acoustic scattering; Blood flow; Calibration; Light scattering; Optical scattering; Particle scattering; Probes; Transducers; Ultrasonic imaging;
  • fLanguage
    English
  • Journal_Title
    Sonics and Ultrasonics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9537
  • Type

    jour

  • DOI
    10.1109/T-SU.1978.31028
  • Filename