• DocumentCode
    1229665
  • Title

    Direct Comparison of Josephson Waveforms Using an AC Quantum Voltmeter

  • Author

    Behr, Ralf ; Palafox, Luis ; Ramm, Günther ; Moser, Harald ; Melcher, Jürgen

  • Author_Institution
    Phys.-Technische Bundesanstalt, Braunschweig
  • Volume
    56
  • Issue
    2
  • fYear
    2007
  • fDate
    4/1/2007 12:00:00 AM
  • Firstpage
    235
  • Lastpage
    238
  • Abstract
    A new method for calibrating ac voltage waveforms based on measuring the difference relative to a synchronously synthesized waveform supplied from a programmable Josephson array is presented. In order to test the accuracy of this approach, the differences between two Josephson waveforms have been measured at the 1.2-V level for frequencies up to 5 kHz. The uncertainties achieved are at least a factor of ten lower than those from the direct sampling method. Below 400 Hz, this new method gives an uncertainty of 5 10-8 (k = 1), while at 1 kHz, the uncertainty is about 2 107 (k = 1)
  • Keywords
    Josephson effect; calibration; voltage measurement; voltmeters; 1 kHz; 1.2 V; AC quantum voltmeter; AC voltage waveforms calibration; Josephson waveforms; programmable Josephson array; waveform synthesis; AC generators; Analog-digital conversion; Detectors; Frequency measurement; Pulse generation; Sampling methods; Signal synthesis; Testing; Voltage measurement; Voltmeters; Josephson arrays; synchronous sampling procedure; waveform synthesis;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2007.891076
  • Filename
    4126829