DocumentCode
1229745
Title
Rise-Time Calibration of 50-GHz Sampling Oscilloscopes: Intercomparison Between PTB and NPL
Author
Bieler, Mark ; Seitz, Steffen ; Spitzer, Meinhard ; Hein, Günter ; Pierz, Klaus ; Siegner, Uwe ; Basu, Anjan M. ; Smith, Andrew J A ; Harper, Matthew R.
Author_Institution
Phys.-Technische Bundesanstalt, Braunschweig
Volume
56
Issue
2
fYear
2007
fDate
4/1/2007 12:00:00 AM
Firstpage
266
Lastpage
270
Abstract
The paper presents the results of the first intercomparison of the rise-time calibration of a 50-GHz sampling oscilloscope between Physikalisch-Technische Bundesanstalt, Germany, and National Physical Laboratory, U.K. (EUROMET project 641). Rise-times between 6.0 and 6.6 ps were obtained using optoelectronic techniques. Detailed uncertainty budgets are given, which yield typical rise-time uncertainties of 1.2 ps. The absolute values of the degree of equivalence are les 0.3 ps with an expanded uncertainty of 1.0 ps and, thus, validate the experimental techniques
Keywords
calibration; oscilloscopes; 50 GHz; 6.0 to 6.6 ps; EUROMET project 641; Germany; National Physical Laboratory; Physikalisch-Technische Bundesanstalt; U.K.; electro-optic sampling; optoelectronic techniques; rise-time calibration; sampling oscilloscopes; Bandwidth; Calibration; Earth Observing System; Laboratories; Optical pulse generation; Oscilloscopes; Pulse measurements; Sampling methods; Switches; Voltage; 50 GHz; Calibration; electrooptic sampling (EOS); intercomparison; oscilloscope; rise-time;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2007.890609
Filename
4126836
Link To Document