DocumentCode :
1229961
Title :
Cascade-Connected Attenuators
Author :
Beatty, R.W.
Author_Institution :
Central Radio Propagation Laboratory, National Bureau of Standards, Washington, D.C.
Volume :
38
Issue :
10
fYear :
1950
Firstpage :
1190
Lastpage :
1190
Abstract :
Two or more calibrated attenuators are often connected in series (cascade) in order to obtain a desired reduction in power level. The total attenuation of the combination is usually obtained by adding the attenuation of each unit. In general, this will not give the correct result because of reflections due to mismatch at the attenuator junctions. It is shown how the correct value of attenuation can be obtained in terms of reflection coefficient measurements at the attenuator junctions. A nomogram of mismatch error for two cascade-connected attenuators is given.
Keywords :
Attenuation measurement; Attenuators; Impedance matching; Network address translation; Radio propagation; Reflection; Voltage;
fLanguage :
English
Journal_Title :
Proceedings of the IRE
Publisher :
ieee
ISSN :
0096-8390
Type :
jour
DOI :
10.1109/JRPROC.1950.233427
Filename :
1701112
Link To Document :
بازگشت