DocumentCode :
1230152
Title :
Resistors With Calculable Frequency Dependencies Up to 1 MHz
Author :
Han Jun Kim ; Rae Duk Lee ; Semenov, Yu.P.
Author_Institution :
Korea Res. Inst. of Stand. & Sci., Daejeon
Volume :
56
Issue :
2
fYear :
2007
fDate :
4/1/2007 12:00:00 AM
Firstpage :
453
Lastpage :
457
Abstract :
Resistance standards from 1 Omega to 100 kOmega, with calculable frequency dependencies of up to 1 MHz, have been developed for the calibration of commercial inductance-capacitance-resistance (LCR) meters and impedance analyzers. The resistors are designed on the basis of single bifilar loops. The typical resistance change from dc to 1 MHz is from 200 to 800 muOmega/Omega. According to the measurement results, the frequency dependencies of the resistors are several times lower than the measurement uncertainty of commercial LCR meters
Keywords :
calibration; electric resistance measurement; measurement standards; resistors; 1 to 100000 ohm; calculable frequency dependencies; calculable standards; calibration; capacitors; resistors; Calibration; Capacitance; Capacitors; Coaxial components; Conductors; Electrical resistance measurement; Frequency estimation; Impedance; Resistors; Transmission line theory; Calculable standards; calibration; capacitors; frequency; resistors;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2007.891121
Filename :
4126873
Link To Document :
بازگشت