Title :
Traceability of Laser Frequency Calibrations at MIKES
Author :
Merimaa, Mikko ; Nyholm, K. ; Vainio, M. ; Lassila, Antti
fDate :
4/1/2007 12:00:00 AM
Abstract :
We describe the traceability chain of length measurements at Centre for Metrology and Accreditation (MIKES) from atomic clocks to the lasers of primary interferometers. Crucial part of the traceability chain, an optical frequency comb generator linking optical frequencies to atomic clocks, is described in detail. The frequency comb generator is used in frequency calibrations of iodine-stabilized lasers, which are operated in compliance with the recommendations of the Practical Realization of the Definition of the Meter. Measured absolute frequencies of iodine-stabilized lasers, time records of the measurements, and the respective Allan deviations demonstrate the solid performance of the MIKES laser frequency standards. The results are in good agreement with the recommended values, as well as with the independent characterizations of the measured lasers.
Keywords :
Atom lasers; Atom optics; Atomic clocks; Calibration; Frequency measurement; Length measurement; Metrology; Optical interferometry; Solid lasers; Time measurement; Iodine-stabilized lasers; optical frequency comb; optical frequency measurement; traceability;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2007.891075