Title :
Application of a Pencil Ionization Chamber (0.34 cm
Volume) for
Co Beams: Experimen
Author :
Neves, Lucio P. ; Perini, Ana P. ; Fernandez-Varea, J.M. ; Caldas, L.V.E.
Author_Institution :
Inst. de Pesquisas Energeticas e Nucl., Comissao Nac. de Energia Nucl. (IPEN-CNEN/SP), Sao Paulo, Brazil
Abstract :
Ionization chambers are the most utilized dosimeters for precise measurements of absorbed dose, as required e.g., in radiotherapy. The most common type, for this application, is the Farmer type with a wall made of PMMA and/or graphite. With the aim to test new materials and configurations of ionization chambers, using low-cost and easily-available components, this work presents the study of a device developed at the Calibration Laboratory of IPEN (São Paulo). This dosimeter is a pencil-type cylindrical ionization chamber, with a wall of PVC coated with graphite, for routine use in 60Co beams. Following international recommendations, several characterization tests were conducted, and all achieved results were found to fulfill those recommendations. The influence of the chamber components on its response was also studied using the PENELOPE Monte Carlo code. According to the obtained results, this prototype presents potential usefulness for routine dosimetric procedures in radiotherapy.
Keywords :
Monte Carlo methods; dosimetry; graphite; ionisation chambers; polymers; radiation therapy; 60Co beams; Calibration Laboratory of IPEN; Farmer type dosimeters; PENELOPE Monte Carlo code; PMMA; PVC; Sao Paulo; absorbed dose; chamber components; characterization tests; graphite; international recommendations; pencil-type cylindrical ionization chamber; radiotherapy; routine dosimetric procedures; Connectors; Ionization chambers; Materials; Monte Carlo methods; Semiconductor process modeling; Stability analysis; Standards; $^{60}$Co beams dosimetry; Monte Carlo simulation; ionization chamber; radiotherapy;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2012.2229294