• DocumentCode
    1231152
  • Title

    Phase Detection of the Two-Port FPW Sensor for Biosensing

  • Author

    Chang, Wen-Yang ; Sung, Po-Hsun ; Chu, Chun-Hsun ; Shih, Ching-Jui ; Lin, Yu-Cheng

  • Author_Institution
    Dept. of Eng. Sci., Nat. Cheng Kung Univ., Tainan
  • Volume
    8
  • Issue
    5
  • fYear
    2008
  • fDate
    5/1/2008 12:00:00 AM
  • Firstpage
    501
  • Lastpage
    507
  • Abstract
    This study reports the phase detection of the two-port flexural plate wave (FPW) sensor for designing and integrating the miniature system and provides a comprehensive methodology for portable using in the biosensor applications of severe acute respiratory syndrome coronavirus (SARS-CoV). The miniature system mainly utilizes the concept of the frequency divider that involves a divider, a time-based oscillator and a gate to reduce the high frequency, and the FPW sensor is fabricated using microelectromechanical systems (MEMS) technologies for producing a potable biosensing detector. The results demonstrate that the insertion loss decreased about -1.15% dB/degC, and the phase delay was about 2.05deg/(1000 cP). The phaseshift resolution was about 10 mV per degree, and the original frequency of 4.2 MHz was divided by 100 to reduce the frequency to 42 kHz. The SARS-CoV could be detected via the S protein binds to the human angiotensin-converting enzyme 2 (hACE2) as a functional receptor, which would cause the phase delay due to the combining of the antibody with the antigen. Therefore, the feasibility studies provide the information that phase detection is an appropriate low-cost technology via frequency divider for fabricating of the miniature biosensors.
  • Keywords
    bioMEMS; biochemistry; biomedical electronics; biosensors; diseases; elastic waves; frequency dividers; microorganisms; microsensors; molecular biophysics; oscillators; pneumodynamics; proteins; readout electronics; MEMS technologies; S protein binding; SARS-CoV; acoustic; antibodies; antigen; biosensor applications; frequency 4.2 MHz; frequency 42 kHz; frequency divider; functional receptor; human angiotensin-converting enzyme 2; microelectromechanical systems; miniature biosensors; miniature system design; miniature system integration; phase delay; phase detection; potable biosensing detector; severe acute respiratory syndrome coronavirus; time-based oscillator; two-port FPW sensor fabrication; two-port flexural plate wave sensor; Biosensors; Delay; Detectors; Frequency conversion; Microelectromechanical systems; Micromechanical devices; Oscillators; Phase detection; Sensor systems; Sensor systems and applications; Biosensing; flexural plate wave; integrated system; phase detection; readout concept; severe acute respiratory syndrome coronavirus (SARS-CoV);
  • fLanguage
    English
  • Journal_Title
    Sensors Journal, IEEE
  • Publisher
    ieee
  • ISSN
    1530-437X
  • Type

    jour

  • DOI
    10.1109/JSEN.2008.918728
  • Filename