• DocumentCode
    1231650
  • Title

    Performance of Argon-Xenon Mixtures in a Gas Proportional Scintillation Counter for the 0.1–10 keV X-Ray Region

  • Author

    do Carmo, S.J.C. ; Borges, F.I.G.M. ; Conde, C.A.N.

  • Author_Institution
    Dept. de Fis., Univ. de Coimbra, Coimbra
  • Volume
    56
  • Issue
    2
  • fYear
    2009
  • fDate
    4/1/2009 12:00:00 AM
  • Firstpage
    437
  • Lastpage
    440
  • Abstract
    The performance of a gas proportional scintillation counter filled with either pure Xe or Ar-Xe mixtures is described for the 0.1-10 keV X-ray energy range. It is shown that the spectra tail distortion exhibited by Xe filled gaseous detectors for soft X-rays below 2 keV is reduced if Ar-Xe mixtures are used. The peak-to-valley ratio increases by a factor of about 2 or 3 as the Xe concentration is reduced from 100% to 5%. Moreover, the energy resolutions for such mixtures are similar to the ones for pure Xe.
  • Keywords
    gas scintillation detectors; proportional counters; X-ray region; Xe filled gaseous detectors; argon-xenon mixtures; electron volt energy 0.1 keV to 10 keV; gas proportional scintillation counter; soft X-rays; Electromagnetic wave absorption; Energy resolution; Instruments; Radiation detectors; Scintillation counters; Solid scintillation detectors; Spectroscopy; Tail; X-ray detection; X-ray detectors; Argon detectors; energy resolution; gas detectors; proportional gas scintillation detectors; x-ray detectors; x-ray spectroscopy; xenon detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2009.2013952
  • Filename
    4812303