Abstract :
The performance of a gas proportional scintillation counter filled with either pure Xe or Ar-Xe mixtures is described for the 0.1-10 keV X-ray energy range. It is shown that the spectra tail distortion exhibited by Xe filled gaseous detectors for soft X-rays below 2 keV is reduced if Ar-Xe mixtures are used. The peak-to-valley ratio increases by a factor of about 2 or 3 as the Xe concentration is reduced from 100% to 5%. Moreover, the energy resolutions for such mixtures are similar to the ones for pure Xe.
Keywords :
gas scintillation detectors; proportional counters; X-ray region; Xe filled gaseous detectors; argon-xenon mixtures; electron volt energy 0.1 keV to 10 keV; gas proportional scintillation counter; soft X-rays; Electromagnetic wave absorption; Energy resolution; Instruments; Radiation detectors; Scintillation counters; Solid scintillation detectors; Spectroscopy; Tail; X-ray detection; X-ray detectors; Argon detectors; energy resolution; gas detectors; proportional gas scintillation detectors; x-ray detectors; x-ray spectroscopy; xenon detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2009.2013952