• DocumentCode
    1234047
  • Title

    Condenser Microphone Sensitivity Measurement by Reactance Tube Null Method

  • Author

    Von Gierke, H.E. ; Von Wittern, W.W.

  • Author_Institution
    Aero Medical Laboratory Engineering Division, Air Materiel Command, Dayton, Ohio.
  • Volume
    39
  • Issue
    6
  • fYear
    1951
  • fDate
    6/1/1951 12:00:00 AM
  • Firstpage
    633
  • Lastpage
    635
  • Abstract
    A reactance tube circuit acting as a sinusoidally varying capacity in the frequency range from 20 to 200,000 cps is used to compensate the capacity variations of a condenser microphone in a high-frequency carrier circuit. The microphone is driven as in the electrostatic actuator method by a voltage of the same frequency, that modulates the reactance tube capacity. By adjusting amplitude and phase of the reactance tube modulation, the total capacity change of the circuit becomes zero. The sensitivity and phase lag of the microphone response can be determined from the actuator and modulating voltages when the circuit is properly calibrated. The circuit is advantageous in obtaining an equivalent pressure chamber calibration of condenser microphones up to 200 kc. The null method makes it useful for relatively unsensitive microphones.
  • Keywords
    Actuators; Amplitude modulation; Calibration; Circuits; Filters; Frequency; Microphones; Oscilloscopes; Phase modulation; Voltage;
  • fLanguage
    English
  • Journal_Title
    Proceedings of the IRE
  • Publisher
    ieee
  • ISSN
    0096-8390
  • Type

    jour

  • DOI
    10.1109/JRPROC.1951.233785
  • Filename
    1701571